Soliwave FDR56/FQR56
16 Endress+Hauser
Electrical connection
5.2.2 Supply circuit connection
2 1
1
2
1 2
0000000163
10 Supplycircuitconnection(Connector1)
Supply voltage
85 to 253V AC, 50/60Hz
20 to 60V DC or 20 to 30V AC, 50/60Hz
Power consumption
FQR56:
- 7VA (85 to 253V AC, 50/60Hz)
- 1W (20 to 60V DC) / 1.5VA (20 to 30V AC, 50/60Hz)
FDR56:
- 9VA (85 to 253V AC, 50/60Hz)
- 2.4W (20 to 60V DC) / 4VA (20 to 30V AC, 50/60Hz)
Electrical connection Supply voltage
Connection terminals Terminals 1 - 2
M12 connector Binder series 713/763 Connector 1, contact 1 - 2
Harting connector type HAN8D Connector 1, contact 1 - 2
Connect the power supply according to the device version
The polarity of the supply voltage can be set as required.
Provide overcurrent protection device (max. 10A) for the supply voltage.
In accordance with IEC/EN61010 a suitable circuit breaker must be provided for the measu-
ring device.
The electrical connection with connector is only available for the power supply with 20 to
60V DC or 20 to 30V AC, 50/60Hz (ordering option "E").