Operation Turbimax CUS51D
30 Endress+Hauser
Field of application Range Application Recommended type of
calibration
Primary sludge
extraction, primary
clarification
3 to approx.
50 g/l
Excess sludge [g/l, %TS] Two-point (outside the
process)
> approx.
50 g/l
Digested sludge[g/l, %TS] Two-point (outside the
process)
Aeration basin 0 to 5 g/l Thin sludge [mg/l, g/l] Single-point (in the
process)
2 to 15 g/l Activated sludge [mg/l, g/l]
Excess sludge
[g/l, %Ts]
Two-point (outside the
process)
Sequencing batch
reactors
0 to approx.
50 g/l
Sludge, general [mg/l, g/l, %TS]
For applications with wide dynamic range,
from clear water to high solids content
Single-point (in the
process)
Recirculation pipe 3 to approx.
50 g/l
Excess sludge [g/l, %TS] Two-point (outside the
process)
Waste activated sludge
extraction
3 to approx.
50 g/l
Excess sludge [g/l, %TS] Two-point (outside the
process)
> approx.
50 g/l
Digested sludge [g/l, %TS] Two-point (outside the
process)
Sludge thickener
(primary sludge)
3 to approx.
50 g/l
Excess sludge [g/l, %TS] Two-point (outside the
process)
> approx.
50 g/l
Digested sludge
[g/l, %TS]
Two-point (outside the
process)
Digester inlet 3 to approx.
50 g/l
Excess sludge [g/l, %TS] Two-point (outside the
process)
> approx.
50 g/l
Digested sludge
[g/l, %TS]
Two-point (outside the
process)
Digester outlet (sludge) > 5 g/l Digested sludge [g/l, %TS] Two-point (outside the
process)
3 to
maximum
50 g/l
Excess sludge
[g/l, %TS]
Two-point (outside the
process)
WWTP outlet 0 to 5 g/l Formazine [FNU, NTU], Thin sludge [mg/l,
g/l]
Kaolin
[mg/l, g/l]
Single-point (in the
process)
Sand filter monitoring 0 to 5 g/l Formazine [FNU, NTU], Thin sludge [mg/l,
g/l]
Single-point (in the
process)
Preferred applications are highlighted in bold.
Application: process water
Field of application Range Application Recommended type of
calibration
Silicon dioxide process
water
0 to 5 g/l Formazine [FNU, NTU],
Thin sludge
[mg/l, g/l], Kaolin [mg/l, g/l])
Single-point (in the
process)
Silicon dioxide process
sludges
5 to 100 g/l SiO2 [ppm, g/l] Two-point (outside the
process)
Titanium dioxide
process water
0 to 1 g/l Formazine [FNU, NTU],
Thin sludge
(mg/l, g/l), Kaolin [mg/l, g/l])
Single-point (in the
process)

Other manuals for Endress+Hauser Turbimax CUS51D