Turbimax CUS71D Commissioning
Endress+Hauser 17
Enter the Zero adjust (distance from the waterline to
the sensor diaphragm)
Outside the Blanking zone (above Upper window
limit and below Lower window limit) the permanent
echo signals are blanked as interference signals.
For upper and lower window limit enter the distance to
the waterline.
In the adjacent example the permanent echo signals are
blanked outside the range of 0.3 to 3.3 m.