Turbimax CUS71D Commissioning
Endress+Hauser 21
Range above bottom establishes a zone near the
bottom of tank that permits special handling of a
dominant signal that originates from the tank floor.
This signal will be suppressed for the calculation of the
measuring value.
Enter a value from 0.0 to 1.0 m.
Bottom signal set point limits gain amplification
when the primary signal is a reflection from the tank
bottom.
It prevents over-amplification of the signal in
applications with low density material, or when the
tank is empty.
Enter a value from 0 to 100.
100 gives maximum gain.
In Tracking you can change the interface parameters
Top layer
Measures an interface consisting of light-density
material that is at a higher elevation in the tank.
Lower layer
Measures an interface consisting of denser material
that is nearer the bottom of the tank.