How to resolve a fan sensor failure on Enthermics EC1850BL?
- CChristopher GreenSep 12, 2025
If you encounter a Fan or Fan Sensor Failure in your Enthermics Medical Equipment (fluid warmers only), check to ensure the fan sensor wires are connected to both the fan and the control board. If the error continues after checking the wires, replace the fan.