Division:
20 I 77
enclosure allows an ergonomic sitting posture as well as an optimal view of the display of the
analytical balance. Thanks to the high interior area, the user has more freedom of movement
for pipetting and handling samples.
The performance of ST1 series safety weighing enclosures was tested with surrogate
substances. Corresponding test results can be requested from the manufacturer if required.
(see appendix E)
ST1 safety weighing enclosure structure
The standard system consists of the ST1 safety weighing enclosure, the ventilation kit and a
filter unit with HEPA filter. The HEPA filter used allows the air to be returned to the room or
the in-house exhaust system to be connected via a draught diverter.
1 ST1 safety weighing enclosure
2 Front flap for inserting larger containers or devices
3 Ventilation kit consisting of a flexible ventilation hose that connects the enclosure with
the filtration system.
4 Filter unit with HEPA filter
5 Air outlet to the room