Fault Action Carried out by
go on
Check the voltage connections, plug in
necessary.
unit does not light up
Check the voltage connections, plug in the mains plug if
necessary.
the filter unit
See "Behaviour in case of malfunction".
sounds continuously
(alarm sound does
not stop)
Check that the filter unit is switched on.
Check that there is free air circulation in the area of the
sensor.
Check airflow with smoke test kit to visualise the flow.
Check connections to sensors
Check air flow with an anemometer.
envirosciences
sounds briefly
(alarm tone stops by
itself)
Check that there is free air circulation in the area of the
sensor.
Check whether there are disturbing air flows near the
enclosure due to draughts (air conditioners, doors or
similar).
Check the room pressure in relation to the corridor (in case
the enclosure is connected to the building's ventilation
system).
Measure air flow (check set
envirosciences
limit value
Check that the filter housing is switched on.
Check that there is free air circulation in the area of the
extraction/filter housing.
Check that the ventilation hose is connected correctly.
Check that the collar is attached correctly.
Readjust the motor control of the filter housing .