63 / 70
7 • Maintenance and servicing
80116-941_V24392
2023-03
AUTO START, bipolar Testing and measuring equipment
Test setup
Fig. 7-9
• The test specimen is connected via the VIO 3 electrosurgical unit to the
power supply.
• The MF-U socket is connected to the VIO test box using the bipolar
adapter cable via the bipolar cable.
Erbe Art. No. Description
20196-053 Bipolar cable, international 2Pin 22
20100-034 Adapter cable bipolar
20100-102 VIO Testbox Auto Start/Auto Stop (bipolar
start/stop)
Testbox 20100-102
2
X X
X X
1
BP1 BP2
S1 S2