7 Maintenance and servicing
80116-941_V24392
2023-03
8. Set the test specimen to:
Forceps, bipolar – softCOAG, bipolar – Effect 1 –AUTO START
9. Hold down the S1 button on the VIO test box until the test specimen
is activated. Activation must begin after the test delay that has been
set.
10. Press the S2 button. The test specimen must terminate the activation.
11. Unplug cable.
HF output power,
COAG, bipolar
Testing and measuring equipment
Erbe Art. No. Description
20196-053 Bipolar cable, international 2Pin 22
20100-034 Adapter cable bipolar
HF power meter
20189-350 Dual-pedal footswitch with ReMode