6 Cleaning and Disinfection Instructions
ers2 – 1-Channel ECG Telemetry System 41
201000401000 / Rev 04
6.3 Chest Belt
Recommended cleaning agent and disinfectant: Ecolab Eltra 40 Extra
6.3.1 Cleaning and Disinfection
Put the chest belt into a laundry bag or mesh bag.
Machine-wash the chest strap at 40 °C for at least 20 minutes, using the special detergent
Ecolab Eltra 40 Extra according to the manufacturer's instructions.
Air-dry the wet or dripping-wet chest belt and take care not to overstretch it while wet.
6.4 Receiver
It is recommended to regularly remove dust etc. from the receiver.
6.4.1 Cleaning
Dampen a clean cloth with distilled water.
Wring excess liquid from the cloth.
Wipe the exterior surfaces of the receiver.
Use another clean cloth to dry the surface of the receiver.
CAUTION Risk of infection
Using the chest belt on another patient may cause cross-infections.
The chest belt is intended for use on a single patient only. The product can be reused if repro-
cessed according to the instructions in this manual and if used on the same patient.