10 Electromagnetic Compatibility (EMC) DIN EN 60601-1-2
52 |
52
201000401000 / Rev 04
10 Electromagnetic Compatibility (EMC) DIN EN 60601-1-2
Changes or modifications to this system not expressly approved by ergoline could cause EMC issues
with this or other equipment.
This system is designed to comply with applicable regulations regarding EMC. Its compliance with
these requirements has been verified. It needs to be installed and put into service according to the
EMC information stated as follows.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The ECG system is intended for use in the electromagnetic environment specified below. It is the responsibility of the customer
or user to ensure that the ECG system is used in such an environment.
Emissions Test Compliance Electromagnetic Environment – Guidance
RF emissions to EN 55011 Group 2 The ECG system must emit electromagnetic energy in
order to perform its intended function. These emissions
may cause interference in nearby electronic equipment.
RF emissions (radiated) to EN 55022/55032,
ANSI C63.4
Class B The ECG system is suitable for use in all establishments,
including domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
RF emissions (conducted) to EN 55022/
55032, ANSI C63.4
Class B
Harmonic emissions to EN 61000-3-2 not applicable
Voltage fluctuations/flicker emissions to
EN 61000-3-3
not applicable