Maintenance
68 Optical Switch
Cleaning LT Connectors
Cleaning LT Connectors
The connectors of the RTUe-9110 external switch can be cleaned using a
mechanical cleaner.
To clean a connector using a mechanical cleaner:
1. Insert the mechanical cleaner into the optical adapter, and push the
outer shell into the cleaner.
Note: The cleaner makes a clicking sound that indicates that the cleaning is done.
2. Verify connector surface with a fiber inspection probe (for example,
EXFO’s FIP).
WARNING
Never look directly into a live fiber. It could cause serious eye
damage. Always use a Fiber Inspection Probe.
WARNING
Before cleaning the connectors, you should suspend the tests to
avoid hazardous radiation exposure.