STANDARD EVENT REGISTER
If any of these bits set, and their respective enable bits are set, the Event Summary Bit (ESB) will set in the STATUS BYTE.
PONPower On:
Indicates that the meter was turned off and on since the last communication.
CME Command Error:
Indicates that there was a syntax or spelling error in the command, or the command received is not supported.
EXE Execution Error:
Indicates that the meter detected an error while attempting to execute a command.
DDE Device Dependent Error:
Indicates that the meter did not operate properly due to some internal error.
OPC Operation Complete:
Indicates that all requested operations have been completed.
4-7
Figure 4-3
Standard Event Registers

Questions and Answers: