High Voltage Power Supply
Type: CARRERA
5.5 Assembly, Installation (only for trained personnel)
Assembly and initial start-up must be carried out by skilled
personnel from FERROTEC GmbH or the customer's skilled
personnel who have been especially trained for this work.
Consult the instruction handbook and technical reference
material supplied for assembly and installation of the electron
beam evaporator.
.
DANGER
The installation of the high voltage power supply must be
carried out only by trained, skilled personnel.
5.5.1 Assembly
- The high voltage power supply is designed to be mounted
into a 19“-Rack.
- Please, keep in mind that the unit must be installed in such a
way that there is unobstructed air inlet and outlet for the
operation of the fan of the high-voltage power supply unit
NOTICE
Check the load-bearing capacity of the assembly rack.
To keep the point of gravity of the whole assembly as low as
possible, it is recommended to install the high-voltage power
supply unit as the lowest device of the group.
5.5.2 Electric Connection
Connection conditions
- Connect the power supply line to the Harting plug in
accordance with the supplied drawing (see Appendix).
- Observe the prescribed safety earthing connections
- Make sure there is a differential current switch.