High Voltage Power Supply
Type: CARRERA
5.5.4 Connection of the high-voltage power supply unit
Connect the high-voltage power supply unit in accordance with
the wiring diagram.
The following components are to be connected:
- High-voltage cable (the cable and connectors are part of the
delivery or they are already connected to the filament power
supply)
- Earthing line
- Control line
- Mains supply line
High voltage cable
The high-voltage cable connects the high-voltage power supply
unit with the filament power supply, which is found in the vacuum
device.
1. The high-voltage cable is connected to the high-voltage power
supply unit with the X305 connector.
NOTICE
Be sure to use the supplied high-voltage cable. It is only this
way that you can avoid electromagnetic interference.
2. Secure the high-voltage cable with the protective metal plate.
NOTICE
Optionally, you can connect max. 4 high-voltage cables to the
high-voltage power supply unit. Please, note that in versions of
the high-voltage power supply unit with more high-voltage
outputs all the outputs must be connected. Or the not used
outputs must be protected with blind connectors.