Fisherbrand 50152755 | 9
Installation
Ambient Conditions
Location Requirements
The incubator must only be operated in a location that meets all of the ambient condition
requirements listed below:
Installation location indoors in dry areas free from drafts.
The dust burden may not exceed the contamination category 2 based on IEC61010-1. Using
the incubator in an atmosphere with electrically conductive dust is prohibited.
The minimal distance to adjacent surfaces must be observed on all sides (see Section
“Incubator Dimensions and Clearances” on page 11)
The operating room must be equipped with appropriate ventilation.
Solid, level, fire-proof surface; no flammable materials opposite to the rear panel of the
incubator.
Vibration-proof substructure (floor stand, lab table) capable of bearing the dead weight of the
incubator and its accessories (particularly if two devices are stacked).
The electrical circuitry of the incubator has been designed for an operating height of up to
2000 m above sea level.
Relative humidity up to 80% (maximum; preferably 60-70%), non condensing.
Should condensation exist, wait until the moisture has evaporated completely before
connecting the incubator to a power source and powering up.
The ambient temperature must be within a range of +18 °C to +32 °C (64.4 °F to 89.6 °F).
Avoid direct exposure to sunlight.
Devices that produce excessive amounts of heat must not be placed near the incubator.
Place the incubator on a floor stand (option; to be ordered separately), never on the lab floor.
This prevents the penetration of dust or dirt into the device.
Power line voltage variations must not exceed ±10 % of the nominal voltage.
Transient surges must lie within the range of levels that normally occur in the power supply
system. The impulse withstand voltage based on surge category II of IEC 60364-4-443 shall
be applied at the nominal voltage level.