| Safety Notes
2
|
50152755 Fisherbrand
Chapter 1
Operational Safety Rules
The following rules must be heeded when working with Fisherbrand incubators:
Observe the sample weight limits specified for your Fisherbrand incubator as a whole and its
shelving in particular; see “Technical Data” on page 47.
Do not load the bottom of the interior workspace to avoid the risk of overheating any samples
placed there and to prevent the temperature sensor from being damaged.
Arrange the samples evenly throughout the work space, making sure not to place them too
closely to the interior walls to ensure a uniform temperature distribution.
Do not load your Fisherbrand incubator with substances that exceed the capabilities of the
available lab apparatus and Personal Protection Equipment to provide sufficient degrees of
protection to users and third parties.
Check the door seal every six months for proper sealing performance and possible damage.
Do not process any samples containing hazardous chemical substances that may be
released into the ambient air through defective seals or may cause corrosion or other defects
on parts of the Fisherbrand incubator.
Warranty
Fisher Scientific International, LLC warrants the operational safety and functions of the
Fisherbrand incubators only under the condition that:
the incubator is operated and serviced exclusively in accordance with its intended purpose
and as described in these operating instructions,
the incubator is not modified,
only original spare parts and accessories that have been approved by Fisher Scientific care
used (third-party spares without Fisher Scientific International, LLC approval void the limited
warranty),
inspections and maintenance are performed at the specified intervals,
an operation verification test is performed after each repair activity.
The warranty is valid from the date of delivery of the incubator to the customer.