Fisherbrand 50152755 | 25
Start-up |
Chapter 5
Electrical Requirements to be Met when
Operating the Incubators in the European Union:
The incubators are designed to operate on an AC mains having a maximum system impedance
Zmax = 0.321 Ohms, as measured at the system transfer point according to EN 61000-3-11,
paragraph 6.2.2. The user must ensure that the device is operated on a mains that complies with
this requirement. Where necessary, it is recommended to have the system impedance confirmed
by the power utility company.
NOTE
Condensation
When taking the incubator into operation for the first time allow some time before
switching on for stabilization to avoid condensation forming on live parts.
If a high-voltage test is to be performed on the unit, it must first be heated for around 30
minutes at 75°C.