Fisherbrand 50152755 | 47
Technical Data
The technical data are valid only for an empty device equipped with three shelves, a spray-painted outer
enclosure and a power line voltage 120 V 60 Hz (Table 1) or 230 V 50/60Hz (Table 2). Options may have an
impact on the specified performance.
Table 1. Technical Data - Gravity Convection Incubators 120 V
Parameter Unit
Fisherbrand
Basic 60L
Gravity
Incubator
Fisherbrand
Basic 100L
Gravity
Incubator
Fisherbrand
Basic 180L
Gravity
Incubator
Process
Work space atmosphere
Min.
Max.
K
°C/°F
Ambient tem-
perature plus
7 °C/12.6 °F
75 °C/167 °F
Ambient
tempera-
ture plus
7°C/12.6°F
75 °C/167°F
Ambient
temperature
plus
7°C/12.6°F
75 °C/167 °F
Temperature deviation from set value at
37 °C (98.6 °F), spatial (*)
typical:
max.:
K ± 0.8
± 0.9
± 0.8
± 0.9
± 0.8
± 0.9
Temperature deviation from set value at 37 °C (98.6 °F),
over time (*)
K ±0.3 ±0.3 ±0.3
Heat-up time (work space unoccupied, from 25 °C
(77 °F) to 98% of final temperature (*)
min 80 80 80
Recovery time (work space unoccupied, door open for
30 s, to 98% starting temperature. (*)
typical:
max.:
min 8
10
10
12
12
14
Heat dissipation to environment (at set temperature of
37 °C (98.6 °F) and room temperature of 25 °C/77 °F)
W 21 ±10% 26 ±10% 31 ±10%
(*) all values measured with unoccupied working space and maximum 3 shelves
Overall dimensions
Height mm/in 720/28.3 820/32.3 920/36.2
Width mm/in 530/20.8 640/25.2 640/25.2
Depth mm/in 565/25.2 565/25.2 738/29.1
Overall weight kg/lbs 40/88 51/112 65/143
Loading capacity
Loading capacity per shelf kg/lbs 25/55
Max. overall loading capacity per device kg/lbs 50/110 50/110 75/165