| Technical Data
48
|
50152755 Fisherbrand
Chapter 13
Electrical data
Power consumption W 300 540 720
Maximum current A 2.5 4.5 6.0
Earthing system (e.g. 1/N/PE) 1/N/PE 1/N/PE 1/N/PE
Power line frequency Hz 50/60
Power line voltage +/- 10 % V 120
IP protection system IP 20
Protection class I
Overvoltage category II
Device fusing, building side A 16
Device fusing, on PCB (T 16A H 250 V) A 2 x 16
Environmental conditions
Min. ambient temperature °C/°F 18/65
Max. ambient temperature °C/°F 32/90
Max. humidity in service,
non condensing
% r.F./
% r.H.
80, non condensing
Min. storage temperature °C/°F 20/68
Max. storage temperature °C/°F 60/140
Max. humidity in storage, non condensing % r.F./
% r.H.
90, non condensing
Post-transport acclimation time h 2
Noise level dB(A) no Inherent noise
Pollution degree 2
Site conditions
Maximum altitude above sea level m/y NN 2000/2187
Minimum side clearance mm/in 50/2
Minimum front clearance mm/in 590 / 23.2 690 / 27.2 814 / 32
Minimum back wall clearance mm/in 80/3.2
Minimum bottom clearance mm/in 200/8
Minimum top clearance mm/in 300/12
Table 1. Technical Data - Gravity Convection Incubators 120 V
Parameter Unit
Fisherbrand
Basic 60L
Gravity
Incubator
Fisherbrand
Basic 100L
Gravity
Incubator
Fisherbrand
Basic 180L
Gravity
Incubator