| Product Description
16
|
50152755 Fisherbrand
Chapter 4
Figure 4-2. Fisherbrand Basic 60L/100L/180L Gravity Incubator Rear View
[1] Door bearing
[2] Access port
[3] Electronics compartment
[4] Leveling foot
Safety Devices
The incubators are equipped with the following safety features:
an overheat protection cut-out feature that shuts down the
incubator
completely when
excessive temperatures occur in the workspace;
dual fuses rated at 16 amperes.
Work Space Atmosphere
To ensure undisturbed operation, the ambient temperature in the operating room must be at least
18 °C (64.4 °F).
The heating system uses this temperature threshold to control the ambient temperature plus 7 °C
(12.6 °F) up to the maximum of 75 °C (167 °F) for gravity convection incubators.