Table 2. Technical Data - Gravity Convection Incubators 230 V
Parameter Unit
Fisherbrand
75L Incubator
Fisherbrand
117L Incubator
Fisherbrand
194L Incubator
Process
Work space atmosphere
Min.
Max.
K
°C/°F
Ambient
temperature
plus
7 °C/12.6 °
F
75 °C/167 °
F
Ambient tem-
perature plus
7°C/12.6°F
75 °C/167°F
Ambient
temperature
plus
7°C/12.6°F
75 °C/167 °F
Temperature deviation from set value at
37 °C (98.6 °F), spatial (*)
typical:
max.:
K ± 0.8
± 0.9
± 0.8
± 0.9
± 0.8
± 0.9
Temperature deviation from set value at 37 °C (98.6 °F),
over time (*)
K ±0.3 ±0.3 ±0.3
Heat-up time (work space unoccupied, from 25 °C
(77 °F) to 98% of final temperature (*)
min 80 80 80
Recovery time (work space unoccupied, door open for
30 s, to 98% starting temperature. (*)
typical:
max.:
min 8
10
10
12
12
14
Heat dissipation to environment (at set temperature of
37 °C (98.6 °F) and room temperature of 25 °C/77 °F)
W 21 ±10% 26 ±10% 31 ±10%
(*) all values measured with unoccupied working space and maximum 3 shelves
Overall dimensions
Height mm/in 720/28.3 820/32.3 920/36.2
Width mm/in 530/20.8 640/25.2 640/25.2
Depth mm/in 565/25.2 565/25.2 738/29.1
Overall weight kg/lbs 40/88 51/112 65/143
Loading capacity
Loading capacity per shelf kg/lbs 25/55
Max. overall loading capacity per device kg/lbs 50/110 50/110 75/165
Electrical data
Power consumption W 140 185 345
Maximum current A 0.6 0.8 1.5
Earthing system (e.g. 1/N/PE) 1/N/PE 1/N/PE 1/N/PE
Power line frequency Hz 50/60
Power line voltage +/- 10 % V 230