A Menu Structure FLUXUS G80x
UMFLUXUS_G8V4-6-1EN, 2018-10-10 145
Input of the min. Reynolds number at which the flow is turbulent.
range: 0...25 500 (rounded to hundreds)
0: the default limit of 3 000 will be used
This display will only be indicated if the SuperUser mode is activated and
Profile bounds = user is selected.
request if an additional correction of the flow velocity is to be defined
on: the correction data can be defined
off: without correction of the flow velocity
This display will only be indicated if the SuperUser mode is activated.
Input of the slope for the correction formula.
range: -2.000...+2.000
0.0: without correction
This display will only be indicated if the SuperUser mode is activated and
Calibration = onis selected.
Input of the offset.
range: -12.7…+12.7 cm/s
0: without offset
This display will only be indicated if the SuperUser mode is activated and
Calibration = onis selected.
SYSTEM settings\Gas-Measuring
selection of the displays for the input of the standard conditions for the gas
measurement
This display will only be indicated if Special Funct.\SYSTEM set-
tings\Measuring\Gas-Measuring is activated.
input of the pressure for the local standard conditions
range: 0.70001...1.30000 bar
x
input of the temperature for the local standard conditions
range: -20.0...+90.0 °C
x
SYSTEM settings\Proc. outputs
selection of the displays for the transmitter outputs setting
selection of the output to be installed
SYSTEM settings\Storing
selection of the displays for the storing of measured values in the data logger
setting of the data logger overflow behavior x
selection of the sample mode
• sample: storing and online transmission of the displayed measured value
• average: storing and online transmission of the average of all measured
values of a storage interval
x
INIT-resis-
tant
SYSTEM settings↕
Gas-Measuring
Norm/Std press.
1.01325 bar
SYSTEM settings↕
Proc. outputs
Install Output ↕
Current I1
Storage mode
sample >AVERAGE<