EasyManua.ls Logo

Flexim FLUXUS G800 - Page 21

Flexim FLUXUS G800
161 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Selection of the Measuring Point FLUXUS G80x
UMFLUXUS_G8V4-6-1EN, 2018-10-10 21
disturbance source: T piece
supply line: l 100 D return line: l 20 D
disturbance source: diffuser
supply line: l 60 D return line: l 10 D
disturbance source: valve
supply line: l 80 D return line: l 20 D
disturbance source: reducer
supply line: l 20 D return line: l 10 D
disturbance source: compressor
supply line: l 100 D
Tab. 5.2: Recommended distance from disturbance sources; D – nominal pipe diameter at the measuring point, l – recommended distance
l
l
l
l
l
l
l
l
l

Table of Contents

Related product manuals