Impulse 6000D, 7000DP
Users Manual
38
Transcutaneous Pacemaker Analyzer Specifications (Impulse 7000DP only)
Test Load Selections
Defibrillator Input
Fixed Load ..........................................................50
Accuracy .............................................................±1 %, non-inductive (<2 μH)
Power Rating.......................................................10 defib pulses of 360 J every 5 minutes
Pacemaker Input
Variable Load......................................................50 to 1500 Ω in 50 Ω steps
Accuracy .............................................................±1 %, non-inductive (<2 μH)
Power Rating.......................................................5 W (average), 40 W (peak) @ 1000
Measurements
Manufacturer Specific Algorithms (plus general purpose default algorithm selection)
GE Responder (1500 & 1700)
MDE 300 (Medical Data Electronics)
Medtronic ERS/Physio Control LIFEPAK
MRL (Medical Research Laboratory/Welch Allyn)
Philips/Agilent/HP
Schiller Medical
ZOLL Medical
(plus a general purpose default algorithm selection)
Current
Range..................................................................4.00 to 250 mA
Accuracy .............................................................±(1% of reading + 0.02 mA)

Table of Contents

Other manuals for Fluke Biomedical IMPULSE 7000DP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke Biomedical IMPULSE 7000DP and is the answer not in the manual?

Fluke Biomedical IMPULSE 7000DP Specifications

General IconGeneral
BrandFluke Biomedical
ModelIMPULSE 7000DP
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals