Impulse 6000D, 7000DP
Users Manual
64
Table Synchronization Time Test Measurements (cont)
Measurement Unit of Measure Description
Pulse Width 50% milliseconds The width of the pulse at 50% of its peak
Pulse Width 10% milliseconds The width of the pulse at 10% of its peak
Table 10. Synchronization Time Test Custom Parameters
Parameter Description
ECG Waveform The ECG waveform the Analyzer should simulate during the test
Preset Energy The energy (in Joules) used for the test
The field user is prompted to set the defibrillator to this value.
Using Pacemaker Test Elements
(Impulse 7000DP only)
Pacemaker tests confirm the basic operation of external
transcutaneous pacemakers by measuring various
pacemaker outputs and timing. These tests do not
operate with the Impulse 6000D analyzer.
Pacer Parameter Test
The Pacer Parameter test provides data about the
accuracy of a pacer’s output. Tables
11 and 12 list the Pacer Parameter test measurements
and custom parameters.