Defibrillator/Transcutaneous Pacemaker Analyzer
Using Pacemaker Test Elements (Impulse 7000DP only)
65
Table 11. Pacer Parameter Test Measurements
Measurement Unit of Measure Description
Pacer Rate Pulses/minute Number of pacer pulses detected by the Analyzer during the test
Pulse Amplitude milliamperes Peak current detected during the test
Pulse Width milliseconds Pacer pulse width as measured by the Analyzer
Pacer Energy millijoules Pacemaker energy output
Table 12. Pacer Parameter Test Custom Parameters
Parameter Description
Input Jacks Specifies where the pacer leads are attached on the Impulse 7000DP
Choose between pacer jacks or defib jacks.
Pacer Load Defines load used for the test
If defib jacks are used for the test, a 50 Ω load is used.
Brand The brand of defibrillator/pacer being tested can be specified to optimize the accuracy of the test.
DUT Rate Expected pacer rate; field user prompted to set pacemaker to this rate
DUT Amplitude Expected pacer amplitude; field user prompted to set the pacemaker output current to this value

Table of Contents

Other manuals for Fluke Biomedical IMPULSE 7000DP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke Biomedical IMPULSE 7000DP and is the answer not in the manual?

Fluke Biomedical IMPULSE 7000DP Specifications

General IconGeneral
BrandFluke Biomedical
ModelIMPULSE 7000DP
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals