Impulse 6000D, 7000DP
Users Manual
76
Table 24. Battery Capacity Test Measurements
Measurement Unit of Measure Description
Charge Time seconds How long it took for the defibrillator to charge to the preset energy specified
The test stores multiple charge times – one per discharge.
Table 25. Battery Capacity Test Custom Parameters
Parameter Description
Preset Energy The energy (in Joules) used for the test
The field user is prompted to set the defibrillator to this value.
Discharges/Minute The number of discharges to be completed each minute
The default value of 1 should normally be used.
Total Discharges The number of discharges to be completed during the test

Table of Contents