Fluke 123/124
Users Manual
48
c
Open the SCOPE INPUTS menu.
d
Open the PROBES submenu.
e
Highlight PROBE AC ADJUST.
f
Open the PROBE AC ADJUST
submenu.
g
Highlight ADJUST 10:1 PROBE.
h
A square wave appears on the
screen.
i
Adjust the trimmer screw in the
probe housing to give an optimum
square wave.
j
Return to normal mode.