Digital Communication Interface
Serial Commands - Alphabetic Listing
For query commands, specifying the MIN, MAX, or DEF parameter causes
the instrument to respond with the minimum, maximum, or default setting
respectively.
For set commands, specifying the MIN, MAX, or DEF parameters causes the
instrument to use the minimum, maximum, or default setting respectively.
‘|’ indicates alternate parameter values.
<n> indicates a number is required.
<prb> indicates a probe number (1 or 2) is required.
<num> indicates numeric value is required.
<bool> indicates a Boolean value (0 or 1) is required. The mnemonics OFF and
ON are also accepted for 0 and 1, respectively.
<conv> indicates a conversion mnemonic is required.
<param> indicates a parameter name is required.
<seri> indicates a serial number is required.
<unit> indicates a temperature unit is required.
<temp> indicates a temperature °C/F is required.
<pass> indicates a password is required.
<port> indicates a port number is required.
<label> indicates an eight character label is required.
<year> indicates a four digit number is required.
<month> indicates a one or two digit number is required.
<day> indicates a one or two digit number is required.
<hour> indicates a one or two digit number is required.
<minute> indicates a one or two digit number is required.
<second> indicates a one or two digit number is required.
<baud> indicates a valid baud number is required.
*CLS
Clear the event registers and the error queue.
Example: *CLS
This command has no response.
*IDN?
Returns the instrument identication string that indicates the manufacturer, model
number, serial number, and rmware version.