1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
50
Example: *IDN?
Response: FLUKE,1524,23456,1.00
*RST
Set the instrument operating parameters to dened conditions.
Example: *RST
This command has no response.
CALCulate:AVERage:CLEar
Clear the statistical functions for all probes.
Example: CALC:AVER:CLE
CALCulate<prb>:AVERage<n>:DATA?
Returns the value of a statistical calculation for a given probe. The CALCulate sufx,
<prb>, species the probe, 1 or 2. Default to 1 if omitted. The AVERage sufx, <n>,
species the calculation type number as listed in Table 19 on page 73. Defaults to 1 if
omitted.
Example: CALC2:AVER1:DATA?
Response: 0.017
CALCulate:AVERage<n>:TYPE?
Returns the keyword for the specied eld type number. The AVERage sufx, <n>,
species the calculation type number. Default to 1 if omitted.
Example: CLAC:AVER4:TYPE?
Response: STD
CALCulate<prb>:CONVert:DATE:CALibrate?
Returns the last calibration date for the specied probe. The CALCulate sufx, <prb>,
species the probe, 1 or 2. Defaults to 1 if omitted. The response is returned in the
format <year>,<month>,<day>.
Example: CALC1:CONV:DATE:CAL?
Response: 2000,9,22
CALCulate<prb>:CONVert:DATE:CALibrate <year>,<month>,<day>
Set the last calibration date for the specied probe.
The CALCulate sufx, <prb>, species the probe, 1 or 2. Defaults to 1 if omitted.
The <year> parameter is a four-digit number, 2000 to 2099.