EasyManua.ls Logo

Fluke 1523 - Page 59

Fluke 1523
105 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
52
Example: CALC2:CONV:PAR:CAT?
Response: “R0”, “A”, “B”, “C”, “MINOP”, “MAXOP”
CALCulate<prb>:CONVert:PARameter:VALue? <param>
Return the value of the specied conversion parameter for the specied probe.
The CALCulate sufx, <prb>, species the probe, 1 or 2. Defaults to 1 if omitted.
The <param> parameter is the keyword of the parameter as listed in Table 21 on page
74.
Example: CALC2:CONV:PAR:VAL? RTPW
Response: 100.0145
CALCulate<prb>:CONVert:PARameter:VALue <param>,<num>
Set the value of the specied conversion parameter for the specied probe.
The CALCulate sufx, <prb>, species the probe, 1 or 2. Defaults to 1 if omitted.
The <param> parameter is the keyword of the parameter as listed in Table 21 on page
74.
The <num> parameter is the new value with type and range per the list in Table 21 on
page 74.
To activate the new values for the parameters MINOP and MAXOP and update the
display, the command CALC:CONV:UPD must be sent, the probe must be removed
and inserted, or the unit power must be cycled off and on.
Example: CALC2:CONV:PAR:VAL RTPW,100.0145
NOTE: This command is protected, which requires a password to set it.
Error ‘-2003, “Command protected”’ is queued if the password has not been entered
with SYST:PASS:CEN
CALCulate<prb>:CONVert:SNUMber?
Return the probe serial number for the specied probe.
The CALCulate sufx, <prb>, species the probe, 1 or 2. Defaults to 1 if omitted.
Example: CALC2:CONV:SNUM?
Response: A_336C
CALCulate<prb>:CONVert:SNUMber <seri>
Set the probe serial number for the specied probe.
The CALCulate sufx, <prb>, species the probe, 1 or 2. Defaults to 1 if omitted.
Shop for Fluke products online at:
1.877.766.5412
www.MyFlukeStore.com

Table of Contents

Related product manuals