53
Digital Communication Interface
Serial Commands - Alphabetic Listing
The <seri> parameter consists of up to ten characters, limited to upper case letters,
numeric digits, and the underscore ‘_’.
Example: CALC2:CONV:SNUM A-336C
NOTE: This command is protected, which requires a password to set it.
Error ‘-2003, “Command protected”’ is queued if the password has not been entered
with SYST:PASS:CEN
CALCulate<prb>:CONVert:TEST? <res>|<mv>[,rjt]
Return the calculated temperature in Celsius for the giving sensor reading.
The CALCulate sufx, <prb>, species the probe, 1 or 2. Defaults to 1 if omitted.
The parameter <res> is the sensor reading in ohms for PRT or thermistor.
The parameter <mv> is the sensor reading in millivolts for thermocouples, with op-
tional reference junction temperature <rjt> in Celsius, which defaults to 0 if omitted.
Example: CALC2:CONV:TEST? 100.0
Response: 0.0
CALCulate<prb>:CONVert:UPDate
Update the display with the latest conguration data from the probe(s).
All probe data is reread and the display(s) are updated accordingly.
Example: CALC2:CONV:UPD
CALCulate<prb>:CONVert:IDENtity?
Return the probe data structure identity number for the specied probe.
The CALCulate sufx, <prb>, species the probe, 1 or 2. Defaults to 1 if omitted.
Example: CALC2:CONV:IDEN?
Response: 23
CALCulate<prb>:CONVert:IDENtity <num>
Set the probe data structure identity number for the specied probe.
The CALCulate sufx, <prb>, species the probe, 1 or 2. Defaults to 1 if omitted.
The <num> parameter is an integer value, 0 to 255.
Example: CALC2:CONV:IDEN 23
NOTE: This command is protected, which requires a password to set it.
Error ‘-2003, “Command protected”’ is queued if the password has not been entered
with SYST:PASS:CEN