1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
54
CALCulate<prb>:CONVert:VERSion?
Return the probe data structure version number for the specied probe.
The CALCulate sufx, <prb>, species the probe, 1 or 2. Defaults to 1 if omitted.
Example: CALC2:CONV:VERS?
Response: 1
CALCulate<prb>:CONVert:VERSion <num>
Set the probe data structure version number for the specied probe.
The CALCulate sufx, <prb>, species the probe, 1 or 2. Defaults to 1 if omitted.
The <num> parameter is an integer value, 0 to 255.
Example: CALC2:CONV:VERS 1
NOTE: This command is protected, which requires a password to set it.
Error ‘-2003, “Command protected”’ is queued if the password has not been entered
with SYST:PASS:CEN
CALibrate:DEVice:AUTO?
Return the state of the automatic power off feature.
The response is ON or OFF.
Example: CAL:DEV:AUTO?
Response: ON
CALibrate:DEVice:AUTO <val>
Set the state of the automatic power off feature.
The <val> parameter is ON or OFF. When turned on, the timer is set to a default of 30
minutes.
Example: CAL:DEV:AUTO ON
CALibrate<chn>:DEVice:LOCK?
Return the lock status of the channel and the inserted probe if any.
The CALibrate sufx, <chn>, species the channel, 1 or 2. Defaults to 1 if omitted.
If no probe is inserted, the response is 0 if the channel is not locked, otherwise 1.
If a probe is inserted, the response is 0 if both the channel and the probe are not
locked; otherwise 1 if either is locked.
The CALibrate<chn>:DEVice:LOCK:DATA query may be used to obtain details of
the channel and probe locks.
Shop for Fluke products online at:
1.877.766.5412
www.MyFlukeStore.com