1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
62
Error ‘-221,”Settings conict”’ is queued if the data logging storage is corrupt.
Example: LOG:AUT:TIM?
Response: 1
LOGging:AUTomatic:TIMe <rate>
Set the Auto-Log interval.
The <rate> parameter must be one of the following values: AUTO, 1, 2, 5, 10, 30, or
60. Numeric intervals are expressed in seconds. AUTO logs unique values as often as
they are available.
Error ‘-221,”Settings conict”’ is queued if the data logging storage is corrupt, if the
LOG menu is active, or if data logging is active.
Example: LOG:AUT:TIM 1
LOGging:AUTomatic:VALue? <num>
Return the specied Auto-Log entry for a preselected tag number.
The <num> parameter species the Auto-Log entry to return. Specifying 0 returns the
rst (oldest) log entry. Specifying a value greater than the number of logged points
returns the last log entry.
Each data line includes the default or user dened tag name, the channel number, the
primary value and units, and the time and date. When both channels are active, records
for each channel alternate. When only one channel is active, no records are sent for the
inactive channel.
The tag number addressed must have been previously selected with the
LOG:AUT:LAB command.
Error ‘-221,”Settings conict”’ is queued if the data logging storage is corrupt, or if no
tag is selected with LOG:AUT:LAB.
Example: LOG:AUT:VAL? 10
Response:
DATA_25,1,23.220,C,15:19:46.7,2000-05-24
DATA_25,2,23.220,C,15:19:46.7,2000-05-24
LOGging:DEMand:DELete
Delete all entries from the demand log.
Error ‘-221,”Settings conict”’ is queued if the data logging storage is corrupt, or if
the demand recall or demand delete display is active.