1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
64
Example: LOG:DEM:PRIN
Response:
DTEST,1,22.676,C,15:19:42,2000-05-24
DTEST,2,9.960,KO,15:19:44,2000-05-24
DTEST,1,23.220,C,15:19:46,2000-05-24
DTEST,2,23.245,C,15:19:48,2000-05-24
DTEST,1,22.765,C,15:19:52,2000-05-24
DTEST,2,9.962,KO,15:19:54,2000-05-24
DTEST,1,23.087,C,15:19:56,2000-05-24
DTEST,2,23.260,C,15:19:58,2000-05-24
LOGging:DEMand:VALue? <num>
Return the specied demand log entry.
The <num> parameter species the demand log entry to return, 1 to 25. Specifying
0 returns the rst (oldest) log entry. Specifying a value greater than the number of
logged points returns the last log entry.
The example shows the 1524 return data. A single channel, and no time and date, are
shown on the 1523 return data.
Each data line includes the default or user dened demand log description, the channel
number, the primary value and units, and on the 1524 the time and date. When both
channels are active, records for each channel alternate. When only one channel is ac-
tive, no records are sent for the inactive channel.
Error ‘-221,”Settings conict”’ is queued if the data logging storage is corrupt.
Example: LOG:DEM:VAL? 10
Response:
DTEST,1,125.820,C,15:19:46,2000-05-24
DTEST,2,23.220,C,15:19:46,2000-05-24
LOGging:DEMand:STATistic<n>? <num>
Return the specied demand log statistical or auxiliary value for the specied demand
log entry.
The STATistic sufx, <n>, species the demand log statistical type number as listed in
Table 23 on page 76. Defaults to 1 if omitted.
The <num> parameter species the demand log entry to return, 1 to 25. Specifying
0 returns the rst (oldest) log entry. Specifying a value greater than the number of
logged points returns the last log entry.
Shop for Fluke products online at: