65
Digital Communication Interface
Serial Commands - Alphabetic Listing
The example shows the 1524 return data. A single channel, and no time and date, are
shown on the 1523 return data.
Each data line includes the default or user dened demand log description, the channel
number, the requested value and units, and on the 1524 the time and date. When both
channels are active, records for each channel alternate. When only one channel is ac-
tive, no records are sent for the inactive channel.
Error ‘-221,”Settings conict”’ is queued if the data logging storage is corrupt.
Example: LOG:DEM:STAT2? 10
Response:
DTEST,1,803.220,C,15:19:46,2000-05-24
DTEST,2,-157.158,C,15:19:46,2000-05-24
LOGging:LABel<n>:NAME?
Return the user dened tag name of the specied auto logging tag number, or the
demand log user dened name.
If the LABel sufx, <n>, is supplied, it species the auto logging tag number of the
user dened tag name to return, 1 to 25. If no user dened name has been entered for
the tag number, the default tag name DATA_nn is returned.
If no sufx is supplied, the demand log user dened name is returned. If no user
dened name has been entered for the demand log, the default name DEMAND is
returned.
Error ‘-221,”Settings conict”’ is queued if the data logging storage is corrupt.
Example: LOG:LAB18:NAME?
Response: DATA_18
LOGging:LABel<n>:NAME <label>
Set the user dened tag name of the specied auto logging tag number, or the demand
log user dened name.
If the LABel sufx, <n>, is supplied, it species the auto logging tag number of the
new name, 1 to 25.
If no sufx is supplied, the new name is saved for the demand log.
The <name> parameter can be up to 8 characters in length and can include any upper
case letters, numeric digits, and the underscore (‘_’).
Error ‘-221,”Settings conict”’ is queued if the data logging storage is corrupt.