1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
66
Example: LOG:LAB18:NAME BATH
MEASure? <prb>
Return the last measurement for the specied probe, with units according to the
UNIT:TEMP setting if it is a temperature.
The <prb> parameter species the probe, 1 or 2. Defaults to 1 if omitted.
If there is no valid measurement available the response is “0.0,OL”.
Example: MEAS? 1
Response: 0.127
READ? <prb>
Return the last measurement for the specied probe, with units according to the
UNIT:TEMP setting if it is a temperature.
The <prb> parameter species the probe, 1 or 2. Defaults to 1 if omitted.
If there is no valid measurement available the response is “0.0,OL”.
Example: READ? 2
Response: 55.011
SENSe<prb>:DATA:MV?
Return the present direct or thermocouple millivolt reading for the specied probe.
The SENSe sufx, <prb>, species the probe, 1 or 2. Defaults to 1 if omitted.
The response is expressed in millivolts. If there is no valid measurement available the
response is “0.0,OL”.
If the presently selected input type is not millivolts or thermocouple, there will be no
response and error ‘-221,"Settings conict"’ will be queued.
Example: SENS2:DATA:MV?
Response: 7.356895
SENSe<prb>:DATA:OHMS?
Return the present direct, PRT, RJ, or thermistor ohms reading for the specied probe.
The SENSe sufx, <prb>, species the probe, 1 or 2. Defaults to 1 if omitted.
The response is expressed in ohms. For a thermocouple probe, the response is the
reference junction ohms value. If there is no valid measurement available the response
is “0.0,OL”.
If the presently selected input type is millivolts, there will be no response and error
‘-221,"Settings conict"’ will be queued.