67
Digital Communication Interface
Serial Commands - Alphabetic Listing
Example: SENS2:DATA:OHMS?
Response: 100.45
SENSe:DATA:RJ?
Return the present internal or external reference junction temperature reading for the
specied probe.
The response is expressed in degrees Celsius.
If the presently selected input type is not millivolts or thermocouple, there will be no
response and error ‘-221,"Settings conict"’ will be queued.
Example: SENS:DATA:RJ?
Response: 23.568
SENSe:RJ:STATe?
Return the state of the reference junction for a given probe.
The response is OFF (no compensation), ON (automatic internal compensation), or
EXT (external compensation). If EXT is selected, the SENS:RJ:TEMP command must
be used to set the external compensation temperature.
Example: SENS:RJ:STAT?
Response: EXT
SENSe:RJ:TEMPerature?
Return the external reference junction temperature for a given probe.
The response is expressed in degrees Celsius.
Example: SENS:RJ:TEMP?
Response: 22.731
SENSe:RJ:TEMP <val>
Set the external reference junction temperature for a given probe.
The parameter <val> is expressed in degrees Celsius, and may range from -10C to
60C.
Example: SENS:RJ:TEMP 23.123
STATus:MEASure:EVENt?
Read and clear the Measurement Event Register, indicating whether or not a new mea-
surement is available to be read.