1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
68
The bit assignments are as follows:
bit 0: probe 1 new measurement
bits 1 to 7: not used
bit 8: probe 2 new measurement
bits 9 to 15: not used
Example: STAT:MEAS?
Response: 257
STATus:MEASure:CONDition?
Return the Measurement Condition Register, indicating whether or not measuring is
active.
The bit assignments are the same as for STATus:MEASure?
Example: STAT:MEAS:COND?
Response: 257
STATus:MEASure:ENABle?
Return the Measurement Event Enable Register.
The bit assignments are the same as for STATus:MEASure?.
Example: STAT:MEAS:ENAB?
Response: 0
STATus:MEASure:ENABle <num>
Set the Measurement Event Enable Register.
The <num> parameter species the event enable bits which are to be set, 0 to 65535
decimal. The bit assignments are the same as for STATus:MEASure?
Example: STAT:MEAS:ENAB 257
STATus:QUEStionable:EVENt?
Read and clear the Questionable Event Register, indicating any measurement
warnings.
The bit assignments are as follows:
bit 0: probe 1 not inserted
bit 1: probe 1 measurement under range
bit 2: probe 1 measurement over range
bit 3: probe 1 EEPROM fault (bad checksum or
version number is not supported)