69
Digital Communication Interface
Serial Commands - Alphabetic Listing
bit 4: probe 1 locked
bits 5 to 7: not used
bit 8: probe 2 not inserted
bit 9: probe 2 measurement under range
bit 10: probe 2 measurement over range
bit 11: probe 2 EEPROM fault (bad checksum or
version number is not supported)
bit 12: probe 2 locked
bits 13 to 14: not used
bit 15: Data logging EEPROM initialization failure
Example: STAT:QUES?
Response: 256
STATus:QUEStionable:CONDition?
Return the Questionable Condition Register, indicating any measurement warning
conditions.
The bit assignments are the same as for the STATus:QUEStionable?.
Example: STAT:QUES:COND?
Response: 256
STATus:QUEStionable:ENABle?
Return the Questionable Event Enable Register.
The bit assignments are the same as for STATus:QUEStionable?
Example: STAT:QUES:ENAB?
Response: 0
STATus:QUEStionable:ENABle <num>
Set the Questionable Event Enable Register.
The <num> parameter species the event enable bits which are to be set, 0 to 65535
decimal. The bit assignments are the same as for the STATus:QUEStionable?.
Example: STAT:QUES:ENAB 1799
SYSTem:COMMunicate:SERial:BAUD?
Return the present serial port baud rate setting.
The response is the present baud rate setting.