2
Safety Information
A Warning identifies conditions and procedures that are
dangerous to the user. A Caution identifies conditions and
procedures that can cause damage to the Product or the
equipment under test.
 Warning
To prevent possible electrical shock, fire, or
personal injury:
Carefully read all instructions.
Read all safety information before you use the
Product.
Do not alter the Product and use only as
specified, or the protection supplied by the
Product can be compromised.
Do not use the Product around explosive gas,
vapor, or in damp or wet environments.
Before each use, examine the Product. Look for
cracks or missing pieces of the clamp housing
or output cable insulation. Also look for loose or
weakened components. Carefully examine the
insulation around the jaws.
Comply with local and national safety codes.
Use personal protective equipment (approved
rubber gloves, face protection, and flame-
resistant clothes) to prevent shock and arc blast
injury where hazardous live conductors are
exposed.
Use Product-approved measurement category
(CAT), voltage, and amperage rated accessories
(probes, test leads, and adapters) for all
measurements.
Do not use the Product if it is altered or
damaged.
Disable the Product if it is damaged.
Do not use the Product if it operates incorrectly.

Other manuals for Fluke 1630-2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1630-2 and is the answer not in the manual?

Fluke 1630-2 Specifications

General IconGeneral
TypeEarth Ground Clamp Meter
Ground Resistance Range0.025 Ω to 1500 Ω
DisplayLCD
Data LoggingYes
Operating Temperature-10 °C to 50 °C
Storage Temperature-20 °C to 60 °C
Ground Resistance Accuracy±(1.5% + 0.02 Ω)
Jaw Opening32 mm
Power SupplyAA batteries
Safety RatingCAT IV 600 V