Logger/Monitor
Specifications
7
Accuracy at Reference Conditions
Parameter Range
Maximum
Resolution
Intrinsic Accuracy at Reference
Conditions
(% of Reading + % of Range)
Voltage
1000 V 0.1 V ±(0.2 % + 0.01 %)
Current
Direct
Input
Rogowski Mode
15 mV 0.01 mV ±(0.3 % + 0.02 %)
150 mV 0.1 mV ±(0.3 % + 0.02 %)
Clamp Mode
50 mV 0.01 mV ±(0.2 % + 0.02 %)
500 mV 0.1 mV ±(0.2 % + 0.02 %)
1500 A Flex
150 A 0.01 A ±(1 % + 0.02 %)
1500 A 0.1 A ±(1 % + 0.02 %)
3000 A Flexi
300 A 1 A ±(1 % + 0.03 %)
3000 A 10 A ±(1 % + 0.03 %)
6000 A Flexi
600 A 1 A ±(1.5 % + 0.03 %)
6000 A 10 A ±(1.5 % + 0.03 %)
40 A
4 A 1 mA ±(0.7 % + 0.02 %)
40 A 10 mA ±(0.7 % + 0.02 %)
Frequency
42.5 Hz to 69 Hz 0.01 Hz ±0.1 %
Aux Input
[1]
10 Vdc 0.1 mV ±(0.2 % + 0.02 %)
Voltage Min/Max
1000 V 0.1 V ±(1 % + 0.1 %)
Current Min/Max
defined by accessory defined by accessory ±(5 % + 0.2 %)
THD on Voltage
1000 % 0.1 % ±(2.5 % + 0.05 %)
THD on Current
1000 % 0.1 % ±(2.5 % + 0.05 %)
Voltage Harmonic 2 to 50
[2]
1000 % 0.1 % ±(2.5 % + 0.05 %)
Unbalance
[2]
100 % 0.1 % ±0.15 %
[1] 1732/1734/1736/1738 only
[2] 1736/1738 only