28 II Ex
Users Manual
20
Continuity Tests
WCaution
To prevent damage to the Product or to the
equipment under test, disconnect circuit
power and discharge all high-voltage
capacitors before you do a continuity test.
The continuity test has a beeper that sounds when a
circuit is complete. You can do continuity tests and not
have to look at the display.
To do a continuity test, set up the Product as shown in
Figure 4.
Push E to turn the continuity beeper on or off.
The continuity function senses intermittent opens and
shorts that last as little as 1 ms. A brief short causes the
Product to emit a short beep.

Other manuals for Fluke 28 II Ex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 28 II Ex and is the answer not in the manual?

Fluke 28 II Ex Specifications

General IconGeneral
Diode testYes
Product typeDigital multimeter
Product colorBlack, Red
Frequency range0.0005 - 199.99 kHz
DC current range0.0000001 - 10 A
DC voltage range0.0001 - 1000 V
Resistance range0.1 - 50000000 Ω
Measuring categoryCAT IV 600V
Placement supportedHandheld multimeter
Basic accuracy (frequency)0.005%
Basic accuracy (DC current)0.2%
Basic accuracy (DC voltage)0.05%
Temperature measurement range-200 - 1090 °C
Battery typeAAA
Power sourceBattery
Battery technologyAlkaline
Number of batteries supported3
Operating altitude0 - 2000 m
Maximum storage altitude10000 m
Operating relative humidity (H-H)0 - 80 %
Weight and Dimensions IconWeight and Dimensions
Depth198.1 mm
Width100 mm
Height6.35 mm
Weight698.5 g