CAT III 1500V TRMS Clamp Meter
Performance Tests
11
V
AC Voltage
Frequency
30 V @ 10 Hz NA 9.5 Hz 10.6 Hz
30 V @ 55 Hz NA 54.2 Hz 55.8 Hz
30 V @ 450 Hz NA 447.3 Hz 452.7 Hz
V
DC Voltage
[1]
-570 V NA -576.2 V -563.8 V
-300 V NA -303.5 V -296.5 V
-50 V NA -51.0 V -49.0 V
0 NA -0.5 V 0.5 V
50 V NA 49.0 V 51.0 V
300 V NA 296.5 V 303.5 V
570 V NA 563.8 V 576.2 V
V
DC High
Voltage
[1][2]
700 NA 688 V 712 V
-700 NA -712 V -688 V
1450 V NA 1431 V 1469 V
-1450 V NA -1469 V -1431 V
N
DC Voltage
0.000 mV NA -0.5 mV 0.5 mV
490 mV NA 484.6 mV 495.4 mV
-490 mV NA -495.4 mV -484.6 mV
D
Resistance
0 Ω NA -0.5 Ω 0.5 Ω
570 Ω NA 563.8 Ω 576.2 Ω
5700 Ω NA 5638 Ω 5672 Ω
57 kΩ NA 56.38 kΩ 57.62 kΩ
D
Capacitance
10 μFNA9.5 μF 10.6 μF
95 μF NA 93.6 μF 96.5 μF
950 μF NA 936 μF965 μF
DC Amps: Jaw
(with 55xxA
short)
[3]
10.0 A NA 9.4 A 10.7 A
-10.0 A NA -10.7 A -9.3 A
Table 2. Performance Tests (cont.)
Test (Switch
Position)
Calibrator Output
52120A
Range
Readout Limit
Low High