CAT III 1500V TRMS Clamp Meter
Calibration Adjustment
23
A AC Jaw Adjustment Procedure
See Table 13 for only the A ac jaw adjustment steps. When the Product shows SAUE on
the LCD, calibration is complete.
Table 13. A AC Jaw Adjustment
Step Display Calibrator Output Operation
1.
C-26 0 A, 0 Hz
Push
2.
C-27 0.2 A, 50 Hz
Push
3.
C-28 17 A, 50 Hz
Push
4.
C-29 3 A, 50 Hz
Push
5.
C-30 3 A, 100 Hz
Push
6.
C-31 3 A, 200 Hz
Push
7.
C-32 3 A, 300 Hz
Push
8.
C-33 3 A, 400 Hz
Push
9.
C-34 3 A, 440 Hz
Push
10.
C-35 STBY Push I
11.
C-36 STBY Push I
12.
C-37 STBY Push I
13.
C-38 STBY Push I
14.
C-39 STBY Push I
15.
C-40 STBY Push I
16.
C-41 STBY Push I
17.
C-42 STBY Push I
18.
C-26 STBY Push e
19.
Save --- ---