393/393 FC
Calibration Manual
18
V DC/mV DC Adjustment Procedure
See Tab le 5 for both the V dc/mV dc adjustment steps. When the Product shows SAUE on
the LCD, calibration is complete.
V DC Adjustment Procedure
See Table 6 for only the V dc adjustment steps. When the Product shows SAUE on the
LCD, calibration is complete.
Table 5. V DC/mV DC Adjustment
Step Display Calibrator Output Operation
1.
C-08 0 V
Push
2.
C-09 600 V
Push
3.
C-10 0 V
Push
4.
C-11 0.5 V
Push
5.
C-08 STBY Push e
6.
Save --- ---
Table 6. V DC Adjustment
Step Display Calibrator Output Operation
1.
C-08 0 V
Push
2.
C-09 600 V
Push
3.
C-10 STBY Push I
4.
C-11 STBY Push I
5.
C-08 STBY Push e
6.
Save --- ---