393/393 FC
Calibration Manual
22
A AC Jaw/iFlex Adjustment Procedure
See Table 12 for both the A ac jaw/iFlex adjustment steps. When the Product shows SAUE
on the LCD, calibration is complete.
Table 12. A AC Jaw/iFlex Adjustment
Step Display Calibrator Output Operation
1.
C-26 0 A, 0 Hz
Push
2.
C-27 0.2 A, 50 Hz
Push
3.
C-28 17 A, 50 Hz
Push
4.
C-29 3 A, 50 Hz
Push
5.
C-30 3 A, 100 Hz
Push
6.
C-31 3 A, 200 Hz
Push
7.
C-32 3 A, 300 Hz
Push
8.
C-33 3 A, 400 Hz
Push
9.
C-34 3 A, 440 Hz
Push
10.
C-35 0 V, 0 Hz
Push
11.
C-36 60 mV, 50 Hz
Push
12.
C-37 30 mV, 50 Hz
Push
13.
C-38 60 mV, 100 Hz
Push
14.
C-39 120 mV, 200 Hz
Push
15.
C-40 180 mV, 300 Hz
Push
16.
C-41 240 mV, 400 Hz
Push
17.
C-42 300 mV, 500 Hz
Push
18.
C-26 STBY Push e
19.
Save --- ---