CAT III 1500V TRMS Clamp Meter
Calibration Adjustment
17
V AC Adjustment Procedure
See Table 4 for the V ac adjustment steps. When the Product shows SAUE on the LCD,
calibration is complete.
Table 4. V AC Adjustment
Step Display Calibrator Output Operation
1.
C-00 0 V, 0 Hz
Push
2.
C-01 600 V, 50 Hz
Push
3.
C-02 300 V, 50 Hz
Push
4.
C-03 300 V, 100 Hz
Push
5.
C-04 300 V, 200 Hz
Push
6.
C-05 300 V, 300 Hz
Push
7.
C-06 300 V, 400 Hz
Push
8.
C-07 300 V, 500 Hz
Push
9.
C-00 STBY Push e
10.
Save --- ---