CAT III 1500V TRMS Clamp Meter
Calibration Adjustment
21
Capacitance Adjustment Procedure
See Table 10 for only the capacitance adjustment steps. When the Product shows SAUE
on the LCD, calibration is complete.
A DC Adjustment Procedure
See Table 11 for the A dc adjustment steps. When the Product shows SAUE on the LCD,
calibration is complete.
Table 10. Capacitance Adjustment
Step Display Calibrator Output Operation
1.
C-12 STBY Push I
2.
C-13 STBY Push I
3.
C-14 STBY Push I
4.
C-15 STBY Push I
5.
C-16 STBY Push I
6.
C-17 STBY Push I
7.
C-18 0.1 μF
Push
8.
C-19 0.5 μF
Push
9.
C-20 1.5 μF
Push
10.
C-21 110 μF
Push
11.
C-22 500 μF
Push
12.
C-23 1000 μF
Push
13.
C-12 STBY Push e
14.
Save --- ---
Table 11. A DC Adjustment
Step Display Calibrator Output Operation
1.
C-24 0 A
Push
2.
C-25 10 A
Push
3.
C-24 STBY Push e
4.
Save --- ---